8800 Series Trace Moisture Analyzer Instruction Manual Appendices
54
The printer should have at least 80 character columns, and it is best to use continuous form paper.
“?” command
Requests a single report of the measurement in the currently selected format.
send:?
C
R
The unit will respond in the format selected with the Data command, for example:
receive:
B
L
-59.3degC
S
P
01:23:45
S
P
NoAlrm
C
R
L
F
The default factory setting for the instrument is to respond in numeric format in the user chosen
units. See the section about the DATA command for more details.
Serial Number Command
send: SN
C
R
receive: Ser
S
P
Num
S
P
nn
C
R
L
F
where nn is the instrument serial number
Who Command
send: W[HO]
C
R
receive: Dev
S
P
Type=XDT
S
P
Ver=3.1
C
R
L
F
Reset Command
send: R[ESET]
S
P
Y
C
R
the unit will perform power on reset.
Login Command
send: L[OGIN]
C
R
receive: Level
S
P
0
C
R
L
F