Endosafe
®
- Portable Test System User’s Guide
Copyright 2003 Page 15
PIPTS101-10
5. Enter the Cartridge lot number. The cartridge lot number will be
stored for later use.
6. Enter the Calibration code. The Calibration code will be stored for
later use.
7. Enter product information by either typing directly into the box or by
dragging information from the product database into the dialog box.
8. Enter the dilution in the form of “1:x mL/mL” or “x units/mL” for
concentration.
9. Insert the cartridge.
10. Select the Send Setup button to send information to the PTS
Reader.
11. Once the Add sample and Click start message is displayed dispense
25 L into each well of the test cartridge and click start on the
screen.
12. While in progress the screen will display messages showing the
current progress.