5-7
In the example below, the temperature was decreased by 22
degrees C. This value was rounded to 20 degrees C. According
to Table 5-2, with a starting temperature of 94 degrees C, we
should subtract 53 seconds from the hold time on the DNA
Thermal Cycler, giving us a new hold time of 27 seconds.
Figure 5-3
Sample conversion of hold times for down ramps (DNA Thermal
Cycler to GeneAmp PCR System 9600)
Note
If methods developed on the GeneAmp PCR System 9600 will
be used on the DNA Thermal Cycler, you can also use
Ta bl e 5 - 2
to
convert the hold times. Simply add the indicated times instead of
subtracting them.