Epson C82302/04 (Serial I/F) Manuale Utente Pagina 255

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 254
8-5
2
Using a cotton swab, coat wells D1 and E1 with mineral oil.
Figure 8-1 shows the location of these wells.
Figure 8-1
Location of sample wells E1 and D1.
3
Place the probe tray on the sample block so that the probe tray
notch faces the front of the instrument.
The RTD probe assembly consists of two cones, one of which
measures the temperature of the sample well. The wire is attached
to the cone that does not measure the temperature of the sample
well. This cone is a dummy probe.
4
Place the probe assembly into wells D1 and E1 so that the dummy
probe sits in D1. Thread the probe wire through the notch in the
probe tray (see Figure 8-2). Make sure the probe is connected to
the digital thermometer.
Vedere la pagina 254
1 2 ... 250 251 252 253 254 255 256 257 258 259 260 ... 293 294

Commenti su questo manuale

Nessun commento