Performance Test Report (R&S ZVR, R&S ZVRE, R&S ZVRL) ZVx
1043.0009.50 5.20 E-7
Item No.
Characteristic
Measure
according
to section
Min. value
Actual value
Max. value
Unit
5 Residual FM
1 MHz
9.99 MHz
149.9 MHz
750 MHz
1000 MHz
2000 MHz
3000 MHz
4000 MHz
5.2.1.5
-
-
-
-
-
-
-
-
___________
___________
___________
___________
___________
___________
___________
___________
1
1
2
5
5
10
20
20
Hz
Hz
Hz
Hz
Hz
Hz
Hz
Hz
6
Port 1
test set
passive
Level accuracy
Level: –10 dBm
w. Opt. ZVR-B10: +3 dBm
9 kHz
20 kHz
40 kHz
100 kHz
1 MHz
2.1 MHz
3 MHz
10 MHz
50 MHz
200 MHz
500 MHz
1000 MHz
1500 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
4000 MHz
5.2.1.6
–1
–1
–1
–1
–1
–0.5
–0.5
–0.5
–0.5
–0.5
–0.5
–0.5
–0.5
–0.5
–0.5
–0.5
–0.5
–0.5
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
1
1
1
1
1
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB