Vedere la pagina 269
ZVx Performance Test Report (R&S ZVR, R&S ZVRE, R&S ZVRL)
1043.0009.50 5.35 E-7
Item No.
Characteristics
Measurement
according to
section
Min. value
Actual
value
Max. value
Unit
2
R&S
ZVR 75
R&S
ZVRE 75
test set
passive
Directivity PORT2
9 kHz
40 kHz
1 MHz
100 MHz
500 MHz
1000 MHz
1500 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
4000 MHz
5.2.3.2
25
33
33
33
33
33
33
33
33
33
20
20
_________
_________
_________
_________
_________
_________
_________
_________
_________
_________
_________
_________
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
2
R&S
ZVR 75
R&S
ZVRE 75
test set
active
Directivity PORT1
300 kHz
1 MHz
5 MHz
500 MHz
1000 MHz
1500 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
4000 MHz
5.2.3.2
5
15
33
33
33
33
33
33
33
20
20
_________
_________
_________
_________
_________
_________
_________
_________
_________
_________
_________
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
2
R&S
ZVR 75
R&S
ZVRE 75
test set
active
Directivity PORT2
300 kHz
1 MHz
5 MHz
500 MHz
1000 MHz
1500 MHz
2000 MHz
2500 MHz
3000 MHz
3500 MHz
4000 MHz
5.3.2.2
5
15
33
33
33
33
33
33
33
20
20
_________
_________
_________
_________
_________
_________
_________
_________
_________
_________
_________
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
3
Option
R&S
ZVR–B21
Attenuators
ATT a1
f = 1 MHz
0 dB
10 dB
20 dB
30 dB
40 dB
50 dB
60 dB
70 dB
5.2.3.3
–2
–0.1
–2
–2
–2
–2
–2
–2
_________
_________
_________
_________
_________
_________
_________
_________
2
0.1
2
2
2
2
2
2
dB
dB
dB
dB
dB
dB
dB
dB