Epson Powerspan Specifiche Pagina 339

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 515
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 338
Performance Test Record (ZVM) ZVx
1127.8500.60 5.18 E-2
5.3 Performance Test Record (ZVM)
The indicated uncertainties refer to the proposed test setup / test procedure.
The expanded measurement uncertainty amounts to k = 2 (confidence level of 95%, Gauss distribution).
Additional measurement uncertainties, which are within the users responsibility, have not been taken
into account (e.g. cable attenuation in harmonics measurements).
We recommend performing a measurement uncertainty analysis for the test setup in question so that
the values specified in the data sheet can be reliably checked.
The measurement uncertainties specified for SSB phase noise and matching of inputs are to be regar-
ded as measurement limits.
Table 5-1: Performance Test Record Generator Specifications
Item Characteristic Measure-
ment acc.
to section
Specification
min. value/Hz
Measured
value/Hz
Specification
max.
value/Hz
Uncer-
tainty / Hz
1 Frequency deviation
0.010 GHz
0.100 GHz
0.5 GHz
1.5 GHz
1.998 GHz
3 GHz
5 GHz
8 GHz
15 GHz
20 GHz
5.2.1.1
-40
-400
-2000
-6000
-7992
-12000
-20000
-32000
-60000
-80000
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
+40
+400
+2000
+6000
+7992
+12000
+20000
+32000
+60000
+80000
2.6
25.7
129
386
514
772
1286
2057
3858
5143
Vedere la pagina 338
1 2 ... 334 335 336 337 338 339 340 341 342 343 344 ... 514 515

Commenti su questo manuale

Nessun commento