ZVx Test Sequence (R&S ZVC, R&S ZVCE)
1043.0009.50 5.51 E-7
Measurement:
Set test frequencies according to test report.
Read marker value on the network analyzer taking the SOURCE POWER
(-10 dB) into account.
P
ermissible noise values:
Model 50, 60 Model 51, 52, 61, 62
20 kHz to 200 kHz -70 dBm -64 dBm
200 kHz to 20 MHz -90 dBm -84 dBm
20 MHz to 3 GHz -100 dBm -94 dBm
3 GHz to 4 GHz -90 dBm -84 dBm
4 GHz to 8 GHz -80 dBm -74 dBm
5.5.2.4 Matching Input b1 and Input b2
(Only with Option R&S ZVR-B25, External Measurements)
Test equipment,
test set 50 :
Power Splitter RVZ, Calibration Kit ZV-Z21, Test Port Cable pair ZV-Z11
Test setup:
ZVC
a1
b2
External On
b1
THRU
Kabel
Kabel
RVZ
Block diagram for measuring the matching of input b1. For measurement of input b2,
the connections from RVZ to Input b1/b2 are reversed.
Settings at the network analyzer:
MODE EXTERNAL
SWEEP LOG SWEEP
CENTER Test frequency
MEAS S11/S22
MARKER = Measured frequency