Epson Powerspan Specifiche Pagina 290

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 515
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 289
ZVx Test Sequence (R&S ZVC, R&S ZVCE)
1043.0009.50 5.55 E-7
Permissible deviations:
ATT a1, a2 ATT b1, b2
< 2 dB < 2 dB
5.5.3.4 Crosstalk
Test equipment: 2 N-short-circuits
Test setup: Connect N-short-circuits to PORT1 and PORT2.
Settings at the network analyzer:
PRESET
SWEEP LOG SWEEP
SOURCE POWER Maximum value
IF BANDWIDTH 1 Hz
AVERAGE POINT AVG 5
MEAS RATIO b1/b2, DRIVE PORT 2 (crosstalk port 1)
RATIO b2/b1, DRIVE PORT 1 (crosstalk port 2)
MARKER Marker frequency = test frequency
Reference
measurement:
None
Measurement: Read out marker value on the network analyzer.
Permissible crosstalk values:
Port2 to Port1
Mod. 60 Mod. 50 Mod. 61, 62 Mod. 51, 52 w. opt. ZVR-B10
20 kHz to 200 kHz –90 dB –90 dB –84 dB –84 dB –75 dB
200 kHz to 5 MHz –120 dB –120 dB –114 dB –114 dB –105 dB
5 MHz to 1 GHz –130 dB –125 dB –124 dB –119 dB –110 dB
1 GHz to 3 GHz –120 dB –115 dB –114 dB –109 dB –100 dB
3 GHz to 4 GHz –110 dB –105 dB –104 dB –99 dB –90 dB
4 GHz to 6 GHz –105 dB –100 dB –99 dB –94 dB –90 dB
6 GHz to 8 GHz –100 dB –95 dB –94 dB –89 dB –85 dB
Vedere la pagina 289
1 2 ... 285 286 287 288 289 290 291 292 293 294 295 ... 514 515

Commenti su questo manuale

Nessun commento